P
Precautions
Thermo Scientific
SL 16 / 16R
v
•
The lid emergency release may be used in emergencies only to recover the samples from the
centrifuge, e.g. during a power failure (see section
“Mechanical emergency lid release”
on
page 7-2
).
•
Never use the centrifuge if parts of its cover panels are damaged or missing.
•
Do not touch the electronic components of the centrifuge or alter any electronic or mechanical
components.
•
Please observe the safety instructions.
Please pay particular attention to the following aspects:
•
Location: well-ventilated environment, set-up on a level and rigid surface with adequate load-bearing
capacity.
•
Rotor installation: make sure the rotor is locked properly into place before operating the
centrifuge.
•
Especially when working with corrosive samples (salt solutions, acids, bases), the accessory parts and
vessel have to be cleaned carefully.
•
Always balance the samples.
Centrifuging hazardous substances:
•
Do not centrifuge explosive or flammable materials or substances which could react violently with
one another.
•
The centrifuge is neither inert nor protected against explosion. Never use the centrifuge in an
explosion-prone environment.
•
Do not centrifuge inflammable substances.
•
Do not centrifuge toxic or radioactive materials or any pathogenic micro-organisms without
suitable safety precautions.
When centrifuging microbiological samples from the Risk Group II (according to the "Laboratory
Bio-safety Manual" of the World Health Organization WHO), aerosol-tight biological seals have
to be used.
For materials in a higher risk group, extra safety measures have to be taken.
•
If toxins or pathogenic substances have gotten into the centrifuge or its parts, appropriate
disinfection measures have to be taken (see
“Disinfection”
on
page 6-3
).
•
Highly corrosive substances which can cause material damage and impair the mechanical stability
of the rotor, should only be centrifuged in corresponding protective tubes.
IF A HAZARDOUS SITUATION OCCURS, TURN OFF THE POWER SUPPLY TO THE
CENTRIFUGE AND LEAVE THE AREA IMMEDIATELY.
Normal use
The centrifuge is to be used for separating materials of different density or particle size suspended in a
liquid.
Maximum sample density at maximum speed: 1,2
g
cm
3
-----------
Summary of Contents for Thermo Scientific SL16
Page 3: ......
Page 21: ...Thermo Scientific SL 16 16R 3 1 3 Control panel Control panel on page 3 2 on page 3 3 ...
Page 24: ...3 Control panel Control panel 3 4 SL 16 16R Thermo Scientific ...
Page 34: ...4 Operation Switching the centrifuge off 4 10 SL 16 16R Thermo Scientific ...
Page 42: ...6 Maintenance and care Service of Thermo Fisher Scientific 6 6 SL 16 16R Thermo Scientific ...
Page 56: ...B Warranty B 2 SL 16 16R Thermo Scientific ...
Page 61: ......