7
Troubleshooting
Troubleshooting by user
Thermo Scientific
SL 16 / 16R
7-5
E-29
The centrifuge cannot be operated.
The run will not be started.
Is the rotor properly installed?
Is it easy to turn the rotor when the lid is open?
Does the rotor rub against the device?
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-30
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-33
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Overpressure in the refrigeration unit.
Clean the air inlet for the condenser.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-34-E-36
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-40
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The centrifuge accelerates too slowly.
Is the rotor properly installed?
Check wether you selected the right bucket.
Is it easy to turn the rotor when the lid is open?
Does the rotor rub against the device?
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-41-E-74
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Failure message
Problem with centrifuge
Possible causes and cures
Summary of Contents for Thermo Scientific SL16
Page 3: ......
Page 21: ...Thermo Scientific SL 16 16R 3 1 3 Control panel Control panel on page 3 2 on page 3 3 ...
Page 24: ...3 Control panel Control panel 3 4 SL 16 16R Thermo Scientific ...
Page 34: ...4 Operation Switching the centrifuge off 4 10 SL 16 16R Thermo Scientific ...
Page 42: ...6 Maintenance and care Service of Thermo Fisher Scientific 6 6 SL 16 16R Thermo Scientific ...
Page 56: ...B Warranty B 2 SL 16 16R Thermo Scientific ...
Page 61: ......