2
Before use
Before setting up
2-2
SL 16 / 16R
Thermo Scientific
Before setting up
1. Check the centrifuge and the packaging for any shipping damage.
Inform the shipping company and Thermo Fisher Scientific immediately if any damage is
discovered.
2. Remove the packaging.
3. Check the order for completeness (see
“Scope of supply”
on
page iii
.).
If the order is incomplete, please contact Thermo Fisher Scientific.
Transporting the centrifuge
•
Due to its weight (see
“Technical Data”
on
page 1-3
), the centrifuge should be carried by several
people.
•
Always lift the centrifuge at both sides.
Figure 2-1.
Lifting the centrifuge at both sides.
•
The centrifuge can be damaged by impacts.
•
Transport the centrifuge upright and if at all possible in its packaging.
Location
The centrifuge should only be operated indoors.
The set-up location must fulfil the following requirements:
•
A safety zone of at least 30 cm must be maintained around the centrifuge.
People and hazardous substances must be kept out of the safety zone while centrifuging.
WARNING
Always lift the centrifuge on both sides.
Never lift the centrifuge by its front or the back panel.
Always remove the rotor before moving the centrifuge.
Summary of Contents for Thermo Scientific SL16
Page 3: ......
Page 21: ...Thermo Scientific SL 16 16R 3 1 3 Control panel Control panel on page 3 2 on page 3 3 ...
Page 24: ...3 Control panel Control panel 3 4 SL 16 16R Thermo Scientific ...
Page 34: ...4 Operation Switching the centrifuge off 4 10 SL 16 16R Thermo Scientific ...
Page 42: ...6 Maintenance and care Service of Thermo Fisher Scientific 6 6 SL 16 16R Thermo Scientific ...
Page 56: ...B Warranty B 2 SL 16 16R Thermo Scientific ...
Page 61: ......