HyPerforma DS 300 Single‑Use Mixer User's Guide
|
45
Table 4.4. Minimum open‑top mixing for adjustable motor head S.U.M.s with standard
impeller sizes.
Left-
to-right
offset
Front-
to-back
position
Lift height
Motor angle
(fixed or
adjustable
motor head
models)
Motor
rpm
Minimum
working
liquid
volume
Turn-
down
ratio
Minimum
hold-up
Maximum
liquid fill
50 L
drum
≈ 1 in. left
≈ 1/2 in.
0
5
°
≈ 180
9.2 L
5.4:1
100 mL
50 L
100 L
drum
≈ 1 in. left
≈ 1/2 in.
0
5
°
≈ 205
9.8 L
10:1
100 mL
100 L
100 L
drum
≈ 1 in. left
≈ 1/2 in.
≈ 7/8 in.
5
°
≈ 140
16.7 L
6:1
100 mL
100 L
200 L
drum
≈ 3/4 in.
left
≈ 1 1/4 in.
≈ 11 1/16 in.
5
°
≈ 150
20 L
10:1
100 mL
200 L
200 L
drum
≈ 3/4 in.
left
≈ 1 1/4 in.
≈ 11 1/8 in.
5
°
≈ 140
22.5 L
8.8:1
100 mL
200 L
300 L
drum
NA due to
flat bottom
NA
≈ 15.24 in.
5
°
≈ 164
61 L
5:1
500 mL
300 L
300 L
drum
NA due to
flat bottom
NA
≈ 15.24 in.
5
°
≈ 124
65 L
5:1
500 mL
300 L
Table 4.5. Country‑specific electrical plug part numbers.
Country
Cat. no.
Europe
SV50177.345
UK
SV50177.346
China
SV50177.347
US (110 VAC)
SV50177.348
US (208 VAC)
SV50177.349
India
SV50177.350
Italy, Chile
SV50177.351
Brazil, Switzerland
SV50177.352
4.4. Part numbers for components, parts, and
accessories
Thermo Scientific
Chapter 4
|