Maintenance and care
Disinfection
6-4
Sorvall
ST 40
Thermo Fisher Scientific
Disinfection
Disinfect the centrifuge immediately whenever infectious material has
spilled during centrifuging.
Use a sprayer whenever possible so that all surfaces are covered evenly.
The rotor chamber and the rotor should be treated preferably with a
neutral disinfectant. A disinfectant spray would be most suitable for this
purpose so that the rotor and accessory surfaces are covered evenly.
Contact the service department of Thermo Fisher Scientific for
questions regarding the use of other disinfectants!
Disinfect the rotor and accessories as follows:
1. Turn off the centrifuge.
2. Pull out the power supply plug.
3. Grab the rotor handle with both hands and press against the green
key.Autolock
TM
At the same time, pull the rotor directly upwards
with both hands and remove it from the motor shaft. Make sure not
to jam the rotor while doing this.
4. Remove the centrifuge tubes and adapters and dispose of them or
disinfect them.
5. Treat the rotor and rotor lid according to the instructions for the
disinfectant (spray or soak in solution). Adhere strictly to the given
application times!
WARNING
Infectious material can get into the centrifuge when a tube breaks or as a result of spills.
Keep in mind the risk of infection when touching articles and take all necessary
precautions!
In case of contamination, make sure that others are not put at risk.
Decontaminate the affected parts immediately.
Take other precautions if need be.
WARNING
Observe the safety precautions and handling instructions for the cleaning agents used.
Summary of Contents for Sorvall ST40
Page 1: ...50119928 1 Thermo Fisher Scientific Sorvall ST 40 Sorvall ST 40R Instruction manual ...
Page 2: ......
Page 3: ...PN 50119928 1 Print 03 09 Instruction manual Sorvall ST 40 Sorvall ST 40R ...
Page 6: ......
Page 10: ...Tables iv Sorvall ST 40 Thermo Fisher Scientific ...
Page 12: ...Figures vi Sorvall ST 40 Thermo Fisher Scientific ...
Page 20: ...xiv Sorvall ST 40 Thermo Fisher Scientific ...
Page 28: ...Introduction and Description Rotor selection 1 8 Sorvall ST 40 Thermo Fisher Scientific ...
Page 36: ...Before use Storage 2 8 Sorvall ST 40 Thermo Fisher Scientific ...
Page 58: ...Operation Switching the centrifuge off 4 18 Sorvall ST 40 Thermo Fisher Scientific ...
Page 80: ...Troubleshooting When to contact customer service 7 8 Sorvall ST 40 Thermo Fisher Scientific ...
Page 88: ...Warranty B 2 Sorvall ST 40 Thermo Fisher Scientific ...
Page 93: ......