Troubleshooting
Troubleshooting by user
Thermo Fisher Scientific
Sorvall
ST 40
7-5
Imbalanced load
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Imbalance detected.
Check the load placed in the rotor.
Check that the rotor cross bolts are well greased.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Check Set Speed
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
The set point speed is higher than the maximum rotor speed.
Correct the value.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-01 - E-12
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-13
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
The check sum in the data memory is incorrect.
The software corrects errors automatically.
Check the values of the set point settings, etc.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-15 - E-16
Temperature sensor broken / controller
defective
Malfunction in the temperature detection.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-17
Speed for rotor detection exceeded
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-21 - E-22
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
The rotor could not be identified.
Check to see if the rotor is properly installed.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-23
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
The speed control measurement returned a different result.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-25-E-27
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Time has expired for the lid lock drive while opening the lid.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-28
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Error during the self-test of the centrifuge program and the
electronics.
Restart the centrifuge.
If an error message appears again, inform Customer Service.
E-29
The centrifuge cannot be operated.
The run will not be started.
Is the rotor properly installed?
Is it easy to turn the rotor when the lid is open?
Does the rotor rub against the device?
Restart the centrifuge.
If an error message appears again, inform Customer Service.
Failure message
Problem with centrifuge
Possible causes and cures
Summary of Contents for Sorvall ST40
Page 1: ...50119928 1 Thermo Fisher Scientific Sorvall ST 40 Sorvall ST 40R Instruction manual ...
Page 2: ......
Page 3: ...PN 50119928 1 Print 03 09 Instruction manual Sorvall ST 40 Sorvall ST 40R ...
Page 6: ......
Page 10: ...Tables iv Sorvall ST 40 Thermo Fisher Scientific ...
Page 12: ...Figures vi Sorvall ST 40 Thermo Fisher Scientific ...
Page 20: ...xiv Sorvall ST 40 Thermo Fisher Scientific ...
Page 28: ...Introduction and Description Rotor selection 1 8 Sorvall ST 40 Thermo Fisher Scientific ...
Page 36: ...Before use Storage 2 8 Sorvall ST 40 Thermo Fisher Scientific ...
Page 58: ...Operation Switching the centrifuge off 4 18 Sorvall ST 40 Thermo Fisher Scientific ...
Page 80: ...Troubleshooting When to contact customer service 7 8 Sorvall ST 40 Thermo Fisher Scientific ...
Page 88: ...Warranty B 2 Sorvall ST 40 Thermo Fisher Scientific ...
Page 93: ......