Maintenance and care
Autoclaving
6-10
Sorvall
ST 40
Thermo Fisher Scientific
Autoclaving
Rotors and adapters can be autoclaved at 121 °C.
The maximum permissible autoclave cycle is 20 minutes at 121 °C.
Clean the rotor before autoclaving and rinse it with distilled water.
Remove all accessories (tubes, adapters) from the rotor. Place the rotor
on a flat surface.
NOTE:
No chemical additives are permitted in the steam.
CAUTION
Never exceed the permitted temperature and duration when autoclaving.
If the rotor shows signs of corrosion or wear, it must be replaced!