Operation
Rotor installation
Thermo Fisher Scientific
Sorvall
ST 40
4-5
Rotor installation
The approved rotors for the
Sorvall
ST 40 are listed in section
“Rotor
selection”
on
page 1-7
. Use only the rotors and accessories from this list
in the centrifuge.
The rotors are equipped with an Autolock
TM
locking system.
This system is used to automatically lock the rotor to the motor shaft.
The rotor does not have to be bolted onto the motor shaft.
Proceed as follows:
1. Open the lid of the centrifuge and if necessary remove any dust,
foreign objects or residue from the chamber.
Autolock
TM
and o-ring must be clean and undamaged.
Figure 4-4.
Autolock
TM
2. Hold the rotor over the shaft and let it slide slowly down the shaft.
The rotor clicks automatically into place.
CAUTION
Unapproved or incorrectly combined accessories can cause serious damage to the
centrifuge!
CAUTION
Do not force the rotor onto the motor shaft.
If the rotor is very light, then it may be necessary to press it onto the shaft with a small
amount of pressure.
Summary of Contents for Sorvall ST40
Page 1: ...50119928 1 Thermo Fisher Scientific Sorvall ST 40 Sorvall ST 40R Instruction manual ...
Page 2: ......
Page 3: ...PN 50119928 1 Print 03 09 Instruction manual Sorvall ST 40 Sorvall ST 40R ...
Page 6: ......
Page 10: ...Tables iv Sorvall ST 40 Thermo Fisher Scientific ...
Page 12: ...Figures vi Sorvall ST 40 Thermo Fisher Scientific ...
Page 20: ...xiv Sorvall ST 40 Thermo Fisher Scientific ...
Page 28: ...Introduction and Description Rotor selection 1 8 Sorvall ST 40 Thermo Fisher Scientific ...
Page 36: ...Before use Storage 2 8 Sorvall ST 40 Thermo Fisher Scientific ...
Page 58: ...Operation Switching the centrifuge off 4 18 Sorvall ST 40 Thermo Fisher Scientific ...
Page 80: ...Troubleshooting When to contact customer service 7 8 Sorvall ST 40 Thermo Fisher Scientific ...
Page 88: ...Warranty B 2 Sorvall ST 40 Thermo Fisher Scientific ...
Page 93: ......