PrintMate A79510100 Issue 2
60
Print is Faint
Error Message / Symptom
Possible Cause
Remedy
Cassette Clamp error.
Open Print Door and print
a Cassette.
Check Cassette is held
securely during printing and
does not move until the
Cassette Clamp releases it.
If movement is seen call the
Thermo Fisher Scientific
Service Department.
Tape jammed.
Open Print Door and Waste
Spool Access.
Use Tape Feed function on
Touch Panel to check that
Tape feeds correctly.
If Tape does not feed
remove and refit Tape (see
Chapter 2 - Changing the
Foil Tape).
Incompatible or old Tape.
Replace Tape with approved spare
(see Appendix A - Spares and
Accessories).
Incompatible Cassette.
Check that Cassette type is included
on Appendix C - Approved Cassette
List.
Print is faint
Instrument Malfunction.
Call the Thermo Fisher Scientific
Service Department.