36 | System Running
Orion 8010cX Ammonia Analyzer
Software Upgrade:
It is used for upgrading software. When
the software is upgraded, insert a USB storage device
containing the software package, and then click the
“Upgrade” button, the system will upgrade automatically, and
the instrument will restart automatically after the system
upgrade.
Service Life of Components:
It is used to read the service
life and expected remaining service life of injection pump
three-way valve, injection pump, ten-way valve and three-way
valve.
Injection Pump Three-way Valve:
It is a component of the
injection pump. Through the service life interface of the
injection pump three-way valve, the overall service life, current
service life and last reset time of the three-way valve can be
seen. When replacing the three-way valve, the three-way
valve needs to be reset.
Injection Pump:
Through the service life interface of the
injection pump, the overall service life, current service life and
last reset time of the injection pump can be seen. When
replacing the three-way valve, the injection pump needs to be
reset.
Ten-way Valve:
Through the service life interface of ten-way
valve, the overall service life, current service life and last reset
time of the ten-way valve can be seen. When replacing the
ten-way valve, the ten-way valve needs to be reset.
Three-way Valve:
Through the service life interface of three-
way valve, the overall service life, current service life and last
reset time of the three-way valve can be seen. When
replacing the three-way valve, the three-way valve needs to
be reset.
Heating Equipment:
It is used to control heating and read
the current temperature of the reaction cell and measuring cell
at the same time. When the heating device is turned off, the
current temperature will gradually decrease. Be sure to turn
on the switch of heating device when exiting the heating
equipment diagnosis, so as to avoid abnormal measurement
problem caused by switching off the heating device.
Input:
Input diagnosis is used to diagnose the state of input.
Through input diagnosis, the state of digital input 1 and digital
input 2 can be seen to be On or Off.
Output:
Output diagnosis is used to diagnose analog outputs
and relays. Set the analog output value, and then an external
instrument can be used to test the corresponding current
value at the same time to compare with the output, so as to
judge whether the analog output is accurate. For a relay, it is
to test whether it is On or Off, set On or Off at the same time,
and then test the state of the relay, so as to know whether the
relay performs according to the setting.