28 | System Running
Orion 8010cX Ammonia Analyzer
Return Key
The navigation bar return key is used to return from the lower
level interface to the previous level interface. The main
interface navigation bar is used to return to the main interface
from all levels of sub-menus.
Parameter Setting
The parameter setting interface is used to set measurement,
calibration, deep cleaning and input/output. If you want to
change the measurement mode, such as setting the manual
mode, automatic mode, measuring range, etc., you need to
enter the Measurement Setting interface. If you want to
change the calibration parameters, including concentration
and measuring range, enter the Calibration Setting interface
for setting. Deep cleaning is used to set the period of deep
cleaning, and the Input/Output Setting is used to set
input/output.
Table 10. Definition of Analysis Parameters
Measurement Parameters Sub-options
Definition of Parameter
Analysis mode
Manual
In this mode, the “Analysis Cycle” cannot be edited, and the
analysis is started only once by pressing [Start] each time. After
the analysis is completed, the equipment enters the idle state.
Automatic
In this mode, the “Analysis Cycle” can be edited. Press [Start]
to start a periodic and continuous analysis. The time interval
between the start of two adjacent analyses is the set analysis
cycle.
Automatic analysis cycle
/
The analysis cycle can be set as a continuous analysis or any
time between 45 minutes and 720 minutes.
Analysis measuring range
Range 1
It is often used to test the water sample within the
concentration range of
0.02 mg/L ~ 2 mg/L.
Range 2
It is often used to test the water sample within the
concentration range of
0.1 mg/L ~ 15 mg/L.
Range 3
It is often used to test the water sample within the
concentration range of
0.5 mg/L ~ 30 mg/L.
Range 4
It is often used to test the water sample within the
concentration range of
2 mg/L ~ 100 mg/L.
Range 5
It is often used to test the water sample within the
concentration range of
5 mg/L ~ 500 mg/L.
Cleaning after analysis
Times of cleaning
Enter the required cleaning times manually.
Note:
When times of cleaning here is set as N, and N times of
cleaning will be added in the automatic analysis process.
Water sample pre-stirring
/
Use an injection pump to extract air, and stir the water sample
in flow cell reversely.