Orion 8010cX Ammonia Analyzer
Installation | 11
Analyzer Mounting Options
Wall Mounting method
Ensure that the firmware on the wall surface has enough load
capacity (at least 4 times the weight of the instrument is
required). The screws on the wall must be strictly selected
and should be suitable for the characteristics of the wall. If the
instrument is damaged due to improper installation, Thermo
Fisher will not assume any responsibility.
For wall mounting, it is recommended to use the
wall-mounting bracket assembly provided in the package. At
first, fix the wall hanging board in the assembly to the wall with
expansion screws, at least 4 expansion screws are needed,
and it is recommended to use more than 6 expansion screws.
Then, take out two hangers from the assembly and fix them at
the corresponding positions on the back of the ammonia
nitrogen detector with M8 screws and gaskets. Finally, the
instrument is secured on already-fixed wall mounting bracket
on the wall surface through a hanging rack already-installed
on back of the instrument. During the installation process, use
a level ruler to ensure that the wall mounting bracket and the
ammonia nitrogen detector are installed to the required level.
Figure 6. Schematic Diagram of Wall Mounted
Ammonia Analyzer