14 | Installation
Orion 8010cX Ammonia Analyzer
c.Wrap PTFE tape around
the threads of the G1/2
stainless steel fitting on
the bottom of the Flow
Cell and screw it into the
electric ball valve. The
specific settings of the
electric ball valve are
described later.
d.Wrap PTFE tape around
the external threads of the
G1/2 to 20 elbow fitting
and screw it into the
internal threads on the
bottom of the electric ball
valve.
e.Cut a section of steel wire
reinforced hose of about
100 mm and insert it into
the extended fitting at the
bottom of the electric ball
valve and insert it as deep
as possible.
f. Insert a T-type equal-
diameter tee fitting at the
other end of the steel wire
reinforced hose and insert
it as deep as possible.
g.Insert the both ends of a
section of steel wire
reinforced hose into the
elbow fitting on the side of
the Flow Cell and the
T-type tee fitting on the
bottom as shown below,
and insert it as deep as
possible.
h.Place the four hose clamps
at the connection of the
T-fitting box steel wire
reinforced hose and tighten
to prevent water leakage.
k. Connect the drain tube to
the T-type equal-diameter
tee fitting. The outlet tube
also uses the steel wire
reinforced hose. The hose
is inserted as deep as
possible, and then it is
locked with a hose clamp
to prevent water leakage.
l. Unscrew the flow cell plug
on the top of the Flow Cell,
insert the PTFE inlet tube
of Ammonia analyzer, and
tighten to prevent the inlet
tube from moving.