3. Select the
Speed
item using the
Down
arrow key and press the
OK
button.
4. Select, for example,
Fast
as the shaking speed using the
Down
arrow
key, and then press the
OK
button.
5. Select the
Time (hh:mm:ss)
item using the
Down
arrow key and press
the
OK
button.
to
6. Select, for example,
20
seconds (=
00:00:20
) as the shaking time by using
the number keys, and then press the
OK
button.
7. Press the
F1
key to accept the selection and return to the
Main
menu.
Programming incubation
You can set the incubation time and temperature in the
Incubation
window.
Note!
The
Incubation
row is only active with the Multiskan FC with incubator model.
Note!
The temperature set in the protocol overrules the startup temperature. If the protocol
does not include incubation, the startup temperature remains.
1. Select the
Incubation
row in the
Main
menu using the
Down
arrow key
and press the
OK
button.
2. Press the
OK
button on the
Temperature
row and use the
Down
arrow
key to select the preferred temperature from the list. The temperature
range is from 25 to 50°C in increments of 1
C. Then press the
OK
button.
3. If
Wait for warm up
is set to
Yes
, the assay will not start until the
temperature is reached. If
Wait for warm up
is set to
No
, the assay will
start.
4. Select the
Time (hh:mm:ss)
item using the
Down
arrow key and press
the
OK
button.
to
5. Select, for example,
5 min
(00:05:00) as the incubation time by using the
number keys, and then press the
OK
button.