![Thermo Scientific Multiskan FC User Manual Download Page 13](http://html.mh-extra.com/html/thermo-scientific/multiskan-fc/multiskan-fc_user-manual_1105193013.webp)
13
Figure 6.
Transport lock bar and tag removed
Installing the filter wheel
1. Unpack the filter wheel (Figure 7). It is delivered in a filter wheel box. Check that all filters are
clean and undamaged.
Figure 7.
Multiskan FC filter wheel
Caution!
Do not touch the filters with your bare hands.
Caution!
When installing the filter wheel, do not touch any other mechanical or electronic part.
2. Place the filter wheel into the filter slot so that the filter wheel numbers (1-8) face outwards (Figure
8). A magnet locking mechanism will automatically lock the wheel in the correct position and the
optical filter position sensor will make sure the correct filter is used during measurement.
Position slot