You can disable/enable samples and sample replicates from the raw data by pressing the
OK
button.
Select
Disable
and press the
OK
button. If the user enables automatically disabled data, the user is
responsible for data quality.
You can print or export the selected data. For further information, refer to Chapter 8: “Viewing
Re
sults”.
Calculated results
The calculated results are shown in the
Interpretations
window.
In situations where the result cannot be calculated from the available data, the result is expressed as
“NaN” (quantitative) or “Undefined" (qualitative). Examples of such situations are:
The divisor is zero (0).
The kinetic calculation cannot be processed.
Concentrations for extrapolated data cannot be calculated when extrapolation is disabled.
The qualitative range is undefined.
If previous calculation results ar
e “NaN”.
Note!
Blank subtraction is made automatically if the plate layout includes blanks. The blank
used in the calculations is always an average of the blanks included in the layout.
Note!
The results of precalculations are not available in the data view. They are only available
in exported or printed reports.
The kinetic timing is relative. This means that each time point is calculated from the first
measurement point with the filter.
You can print or export the selected data. For further informat
ion, refer to Chapter 8: “Viewing
Results”.
Calibration curve
The calibration curve is shown in the
Calibration curve
window.
The report also includes the calculated fit parameters of the curve.