Chapter 7 Configuration and Setup
7-34
┌────────────────────────────────────────┐
│╒
REFERENCE COMPENSATION
═╕
│
││
Mode: Enabled
│
─────
│
││
│
Reference
│
││
Reference : 4.004um
│
Compensation
│
││
│
─────
│
││
Meas Drift: 5.0%
│
│
││
│
1=Change
│
│└───────────────────────┘
=Prev =Next
│
└────────────────────────────────────────┘
Reference Compensation Summary Display
Reference Compensation Mode
Reference Compensation Mode sets this mode as Off or Enable. The last mode condition
configured is shown on the display. If this is the mode condition for the present
application, press
ENTER
. The display changes to the next parameter to be configured.
Press
1
to change the mode condition. The Mode Select Display appears. Select the
condition for the application (Off or Enable). The display responds "-ACCEPTED-" and
then changes to the next parameter to be configured.
Reference Wavelength
Reference Wavelength specifies the reference wavelength. The last reference wavelength
configured is shown on the display. If this is the desired wavelength, press
ENTER
. The
display changes to the next parameter to be configured. Press
1
to change the
wavelength. The Wavelength Enter Display appears. Key in the desired wavelength (in
m) and press
ENTER
. If the entry is valid, the display responds "-ACCEPTED-" and
then changes to the next parameter to be configured.
Reference Measured Drift
Reference Measured Drift specifies the percent change of intensity at the reference
wavelength between the zero and the measurement cycles prior to displaying a reference
drift error. The last reference measured drift configured is shown on the display. If this
is the correct allowed drift, press
ENTER
.
The display changes to the Humidity Compensation Summary. Press
1
to change the
allowed drift. The Reference Measured Drift Enter Display appears. Key in the desired
allowed drift (0.0 to 100.0) and press
ENTER
. If the entry is valid, the display responds "-
ACCEPTED-" and then changes to the Humidity Compensation Summary.
Summary of Contents for MIRAN 205B Series
Page 4: ......
Page 14: ......
Page 40: ......
Page 54: ......
Page 96: ......
Page 102: ......
Page 108: ......
Page 112: ......
Page 118: ......
Page 122: ...Chapter 13 Infrared Spectroscopy Tutorial 13 2 Figure 13 1 Electromagnetic Spectrum ...
Page 142: ......
Page 170: ......
Page 196: ......
Page 202: ......
Page 210: ......