
10-2
Long Term Storage
NiCd batteries discharge at the rate of 30% per month at room temperature. Be sure to
fully charge any battery pack retrieved from storage prior to using it in the analyzer.
Recharging
For a complete description of battery pack operations with either the Model 5025A or
Model 5100A Charger/Power Supply (see Chapter 3, “Charger/Power Supply”).
CARE OF OPTICS
High concentrations of some compounds, examples of which are listed in Table 10-1,
may cause incorrect readings or degrade the analyzer's cell windows. This could lead to
an error message and require replacement of the windows. The problem may be
minimized by:
Avoiding repeated exposure to such concentrations.
Purging the analyzer thoroughly with clean dry air after exposure to such
concentrations.
Table 10-1.
Compounds That May Degrade Cell Windows
Compound Concentration
Ammonia 250
ppm
Diethylamine 100
ppm
Dimethylamine 100
ppm
Ethanolamine 100
ppm
Hydrogen Cyanide
100 ppm
Hydrogen Fluoride
100ppm
Methylamine 100
ppm
Summary of Contents for MIRAN 205B Series
Page 4: ......
Page 14: ......
Page 40: ......
Page 54: ......
Page 96: ......
Page 102: ......
Page 108: ......
Page 112: ......
Page 118: ......
Page 122: ...Chapter 13 Infrared Spectroscopy Tutorial 13 2 Figure 13 1 Electromagnetic Spectrum ...
Page 142: ......
Page 170: ......
Page 196: ......
Page 202: ......
Page 210: ......