Operation
3-28
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
The following provides a brief overview of how to start a monitoring and
sampling run.
Before starting a run, a HEPA filter or optional 37-mm filter must be
inserted downstream of the optical assembly. The recommended 37-mm
filters are glass microfibre filters. If using the 5-micron PVC filters, use the
glass microfibre filters as a backing media when using battery power.
Equipment Damage
At no time should the ADR-1500 be running without
a filter in place, otherwise serious damage to the internal components may
result.
▲
Before initiating a measurement run it is advisable to perform the zeroing
and automatic internal check out sequences, to ensure optimal operation.
●
From the Operate menu, press the
ENTER
key and scroll the
▲
or
▼
key to the Zero Instrument screen.
ZERO INSTRUMENT
FILTER READY
Attach a HEPA filter to the Total Inlet and press
ENTER
. Do not zero
through the cyclone.
During zeroing, each range within the auto-ranging potential is used to
measure the optical background of the instrument and each value is stored
for subsequent data calculations. The duration of a Zero Operation is
usually 2-3 minutes. After Zeroing is complete, the following screen should
be present:
ZERO INSTRUMENT
COMPLETE: BKG OK
●
From the Zero Instrument screen, by pressing the
▼
key, the user
proceeds to a series of logging screens in the following order:
Starting a Run
Filter Requirements
ZERO/Initialize
Operation
Summary of Contents for MIE ADR-1500
Page 1: ...MIE ADR 1500 Instruction Manual Particulate Monitor Part Number 108836 00 15Nov2017 ...
Page 4: ......
Page 14: ...Figures x MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Page 16: ...Tables xii MIE ADR 1500 Instruction Manual Thermo Fisher Scientific ...
Page 26: ......
Page 132: ......
Page 140: ......