Maintenance and Service
5-26
MIE ADR-1500 Instruction Manual
Thermo Fisher Scientific
Use the following procedure to replace the in-line flow meter filter with
fittings (
Figure 5–11
).
1.
Turn the instrument OFF. Remove the AC plug from the bottom of
the instrument.
2.
Unfasten the two captive hardware and swing cover down on the center
panel.
3.
Disconnect the battery cable from J6 or J8 from the power
communication board.
4.
Use needle nose pliers and release the fitting from the tubing.
5.
To install the new filter, follow the previous steps in reverse order.
Note
When replacing the new filter make sure filter arrow is facing
towards the right.
▲
In-Line Flow Meter
Filter with Fittings
Replacement