The two power fuses are located above the main switch [a] on the rear
side of the microtome.
• To replace the two fuses, turn off the main switch [a] of the instru-
ment and unplug it.
• Then open the cover [b] using a flat screwdriver.
• Pull out the fuse holders and replace the fuses with new ones.
• Put back the fuse holder completly and close the cover [b].
For power requirements 100 – 240 V, 2 fuses T2,5AH slow-blow
Replacement
Work
Rating of Fuses
Replacing the Fuses:
[b]
[a]
Maintenance and Care
Replacement Work