To exchange the movable clamping jaw [c], unscrew the clamping screw [a]
to remove the jaw and replace it.
To cut round specimens, the insert for round specimens with defined
diameters of 6, 15, 19 and 25 mm (special sizes on request) or the V-
insert can be clamped into the standard specimen clamp.
• The pin [d], which fits into the lower clamping jaws [c], positions
the insert precisely. The two springs [b] help to remove the speci-
men from the inserts.
• To fasten the V-distance piece on the fixed jaw of the standard
specimen clamp, the clamping screw must be unscrewed from the
spindle.
• Pull the spindle off the clamp.
• After having inserted the V-distance piece, insert the spindle and the
clamping screw again.
Insert for Round Specimens,
V-Insert and
V-Distance Piece