Thermo Fisher Scientific
Operation Manual HM335S
46
Note!
To clamp specimens, different systems are available. With the orienting
adapter it is simple to align the specimen properly to the knife.
The universal cassette clamp allows a quick change system.
• To insert or remove the cassette from between the fixed [b] and
movable [c] jaws, pull the lever [a] to the front (in the direction of
the arrow).
Note!
To achieve optimal clamping keep the locating surface of the cassette free of
paraffin.
• To insert or remove cassettes between the fixed [b] jaw and the
moveable jaw [c], move the lever [a] upwards.
• To adjust the size of a cassette size which is not to the norm, use the
adjustment nut [d].
To cut big specimen, two different Macro-Universal-Cassette clamps
(Macro-UCC, adjustable) are available for use with MacrOflow-
Cassettes or other commercially available Macro-Cassettes.
Specimen Clamping
Universal Cassette
Clamp
Universal Cassette
Clamp, adjustable
[a]
[c]
[d]
[b]
Operating Instructions
Specimen Clamping
[a]
[a]
[b]
[c]
[d]
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com