Operation Manual HM335S
28
The required section and trimming thicknesses are set with the knob [A].
• To choose between section thickness and trimming thickness, press
the knob [A]
In the NORMAL display mode the thickness range is indicated in the
display.
Note!
When switching over from trim sectioning to fine sectioning by using knob
[A], the motorized cutting speed can be reduced. (see part “setting cutting
speed“ page 39)
FEED pre-selected section thickness
TRIM pre-selected trimming thickness
Setting Section and
Trimming Thickness
The graduation of the section thicknesses is divided into five ranges:
Range
Graduation
up to 5 µm
0,5 µm
from 5 µm to 20 µm
1 µm
from 20 µm to 30 µm
2 µm
from 30 µm to 60 µm
5 µm
from 60 µm to 100 µm
10 µm
The graduation of the trimming thicknesses is divided into four ranges:
Range
Graduation
up to 30 µm
5 µm
from 30 µm to 100 µm
10 µm
from 100 µm to 200 µm
20 µm
from 200 µm to 500 µm
50 µm
Operating Instructions
Display and Key Functions