| Main Controller: 1x8 & 5x16 Segment Programmable
6-45
|
304243H02
Mini-mite Tube Furnace
Chapter 6
8. To set the first Dwell, press SCROLL
button to select
‘DWEL.1’
. Press DOWN
or UP
button to set the value.
9. Now repeat the above three steps (6, 7 & 8) to set remaining all segments.
To Operate the Programmer
Programs can also be operated from the
‘P.STAT’
parameter found in the level 1
parameter list.
OR
OR
NOTE
If not all the segments are used for a program, the Ramp & Dwell of each of the
subsequent Segments should be set to OFF.
A program will end in one of two ways, either revert to the control Set Point or
dwell at the temperature set in the last segment used. When a program finishes
on a dwell and the dwell time expires the temperature will revert to the control set
point.
Before running a program ensure that the control set point is set to Zero to avoid
unexpected heating at the end of the program.
When a Holdback is set, each segment used must have a Ramp Rate assigned to
it, in order for it to be recognized by the program.
Operation
Action
Indication
To Run a program
Press and quickly
r
Beacon RUN = On
Scrolling display - CURRENT
PROGRAM STATE
To Hold a program
Press and quickly
r
Beacon RUN = Flashing
Scrolling display - PROGRAM HOLD
To Reset a program
Press and hold
+
for more than 1
second
If program has ended then
‘PROGRAM END’ will be displayed at
the end of the program
Summary of Contents for Lindberg/Blue M Mini-Mite TF55030
Page 11: ... Safety Precautions 1 9 304243H02 Mini mite Tube Furnace Chapter 1 ...
Page 19: ... Installation 4 17 304243H02 Mini mite Tube Furnace Chapter 4 ...
Page 21: ... Start Up 5 19 304243H02 Mini mite Tube Furnace Chapter 5 ...
Page 57: ... Maintenance Cleaning 8 55 304243H02 Mini mite Tube Furnace Chapter 8 ...
Page 61: ... Wiring Diagrams 10 59 304243H02 Mini mite Tube Furnace Chapter 10 ...
Page 67: ... Contact Information Chapter 13 13 65 304243H02 Mini mite Tube Furnace ...