To Configure the Programmer
1. Enter level 2: refer to section “To Enter Level 2” for steps to enter Level 2.
2. To select the Programmer, press
as many times as necessary to view
‘PROG’.
3. To configure the first Program, press DOWN
or UP
to select program
number
‘1’.
Similarly you can configure program (1, 2, 3, 4 or 5) for configuration.
4. To set the ramp unit, press SCROLL
button to select
‘RAMP.U’
and then
press DOWN
or UP
button to select
hour, min or sec
(In this example
the ramp unit is set in min).
5. To set the Dwell unit, press SCROLL
button to select
‘DWEL.U’
and then
Press
or
to select
hour or min
(In this example the dwell unit is set in
min).
6. To set the first Target Set point, press SCROLL
button to select
‘TSP.1’
. Then
Press DOWN
or UP
button to set the value.
7. To set the first Ramp rate, press SCROLL
button to select
‘RMP.1’
. Press
DOWN
or UP
button to set the value.