Mini-mite Tube Furnace
304243H02 |
6-22
Main Controller: 1x8 & 5x16 Segment Programmable |
Chapter 6
Operator Buttons
Single Set Point Operation
3216 controller has capability to select set point-1 and set point-2. User can setup two
different set points to select the desired set point, SP.SEL function can be used.
To set the desired temperature set point, complete the following steps:
1. Press the SCROLL
button until SP1 or SP2 is displayed.
2. Press UP
or DOWN
button until the desired setpoint is displayed and then
release the button. A few seconds after the button is released, the controller will
accept the new value and is indicated by a brief flash of the display.
3. Press PAGE
button to return to the HOME display.
Press to select a new list of parameters and from any
display - press PAGE to return to the HOME Page
Press to select new parameter from the page header. If held
down it will continuously scroll through parameters.
+
Hold down SCROLL and press UP ARROW to scroll back
parameters.
Press to decrease or change the state of a value.
Press to increase or change the state of a value.