It is very important to keep the KingFisher Flex
heads away from each other and other magnets at all times.
Interference of the magnets on one another may cause serious
damage to the magnets.
s
Q5: How strong are the magnets? Can they, for example,
disturb some sensitive equipment?
A5: The KingFisher Flex should not be kept in close proximity
to magnetic tapes, computer discs or other magnetic storage
systems, such as credit cards, as these can be damaged by
the strong magnetic field of the KingFisher Flex heads.
Do not hold the KingFisher Flex heads close to a PC display,
since this may cause damage to the display. Do not use metal
tools when handling KingFisher Flex heads.
Warning
This product contains very strong permanent
magnets. People wearing a pacemaker or metallic prostheses
should not use this product. A pacemaker or prostheses may
be affected or damaged if it comes in close contact with a
strong magnetic field.
s
Q6: How can the KingFisher Flex heads be changed?
A6: By running the Change_magnet protocol under the
Maintenance menu using the up and down cursor keys.
Q7: Can I concentrate the sample during the run?
A7: Both deep well plates and KingFisher 96 plates can be
used during the same run. Therefore, it is possible to start the
processing by using larger volumes (in a deep well plate) and
elute the purified sample to a smaller volume (in a KingFisher
96 plate).
Q8: How does the heating block work?
A8: The heating block is located inside the instrument and can
be used automatically during the protocol. All KingFisher Flex
plates can be heated using specially designed,
interchangeable heating blocks. Any number of heating steps
can be added to the protocol. During the protocol, when the
protocol enters the heating step, the plate is automatically
moved to the dedicated heating position for heating. After the
heating step the protocol continues automatically to the next
step.
Summary of Contents for KingFisher Flex
Page 1: ...Thermo Scientific KingFisher Flex User Manual Rev 1 5 Cat no N07669 ...
Page 90: ...Glossary 90 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific ...
Page 93: ...Thermo Fisher Scientific Thermo Scientific KingFisher Flex User Manual 93 Notes ...
Page 94: ...94 Thermo Scientific KingFisher Flex User Manual Thermo Fisher Scientific Notes ...