Installation
Installation setups
26 Thermo Scientific KingFisher Flex User Manual
Thermo Fisher Scientific
4. Remove the transport lock screw from the label (
Figure 3
–13.
Removing the transport lock of the heating block (B)
5. Then fasten the screw clockwise to its storage site on the heating block
base. The two screws on the heating block base guide the heating
block into the correct position.
Note
Keep the transport locks (
) for
future transportation of the instrument.
s
Both the transport locks have been successfully removed. When
relocating the instrument, refer to
“How to refit the transport lock of the
” on page 35 and “How to refit the transport lock of the tip
1. Connect the mains supply cable to the mains power connector (
) at the bottom left of the back/side panel. If you need to use any
other type of mains supply cable than supplied, use only cables
certified by the local authorities. Before you plug in the power cable,
ensure that the voltage on the type label at the bottom left of the
back/side panel (
) corresponds to the local voltage.
2. Connect the instrument to a correctly installed line power outlet that
has a protective conductor that is grounded.
How to ensure
startup