iCAP Customer Familiarisation Manual
Page 57 of 72
This document is part numbered 849940090071
Version 1.02
Note that any time the 6-port valve body is removed from its actuator; the valve will require
retraining (re-initialization). { XE "valve:retraining" }
To replace or re-orientate the 6-port valve:
Remove all of the tube connectors to the 6 port valve:
Using the provided hex key (9/64), loosen the hex screw on the locking collar which secures the
base of the valve to the instrument:
Firmly but carefully remove the valve head:
Insert the new valve, or reinsert the existing valve, at the desired angle. Rotate it so that the
nebulizer port will be as close as possible to the nebulizer.