iCAP Customer Familiarisation Manual
Page 47 of 72
This document is part numbered 849940090071
Version 1.02
7.12 Shutting Down the System
Follow the procedure detail earlier.
On the Dashboard Page, click on the “green ringed” Qtegra Driven icon.
Figure 5-9. System shut down
For the 7600 Sprint Valve open the Sprint Valve Configurator software from the desktop icon.
When the software starts most of the options will be grayed out, click “Connect to Sprint Valve”.
Expand the screen using the arrow button on the right side of the window to open manual
controls.
Verify both Sample and Rinse probes are in air, verify or set to Load, turn On the Vacuum Pump
Control for several seconds to evacuate the Sample Loop and Valve then turn Off.