iCAP Customer Familiarisation Manual
Page 55 of 72
This document is part numbered 849940090071
Version 1.02
Place one of the screws through the stator and separator and reattach it to the valve, in the
orientation noted during removal:
Using the hexagonal key, tighten all three screws, to firmly set the parts in place:
NOTE
The screws should be finger tight with the separator unable to move independently, over
tightening can cause damage to the rotor.