Handling and Control
Settings
Thermo Scientific
Heratherm 7-11
v
Temperature Display Unit
The
Settings ->°C / °F
menu item allows for toggling the unit used for displaying
temperatures between degrees Centigrade and Fahrenheit.
Note
This setting does not have any impact on data logging via the RS-232 interface. Any
temperature data that is logged to a computer for operational parameter documentation purposes is
handed over in °C.
The word
USEr
appears in the display pane.
Press
to confirm the selection.
In the settings dialog that appears, set the temperature
measured with the external reference sensor by using
or
and confirm your settings with
.
The newly entered value will be stored and used to
calibrate the internal temperature sensors with the
value measured by the reference sensor.
The display returns to its default mode.
The
Settings
icon in the menu bar will go out.
Table 7-11Entering the Calibration Reference Temperature Manually
Table 7-12Toggling the Temperature Display Unit
Press
to activate the menu bar, then use
or
to select the
Settings
icon and press
to
confirm.
Press
to switch to the
C - F
menu item.
The text
C - F
is flashing in the display pane.
Summary of Contents for Heratherm OGS 100
Page 1: ......
Page 6: ...Contents 4 Heratherm Thermo Scientific ...
Page 8: ...List of Figures 2 Heratherm Thermo Scientific ...
Page 34: ...Product Description Work Space Components 4 10 Heratherm Thermo Scientific ...
Page 46: ...Start up Connecting the RS 232 Interface 5 12 Heratherm Thermo Electron LED GmbH ...
Page 62: ...Handling and Control Settings 7 14 Heratherm Thermo Scientific ...
Page 64: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific ...
Page 68: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific ...
Page 84: ...Device Log 15 2 Heratherm Thermo Scientific ...