Product Description
Data Communications & Alarm Interface
Thermo Scientific
Heratherm 4-7
When the user acknowledges the error message, the red alarm icon (D3 in
) is illuminated and the Temperature Set Value icon (see
highlighted by a red border to indicate that thermal protection has kicked in.
Data Communications & Alarm Interface
All signal connections are installed in the electrical interface panel at the rear of the oven.
RS-232 Interface
The RS- 232 interface (left in
below) may be used to connect ovens to the serial
interface port of a computer to allow for the computer-aided acquisition and documentation
of major operating parameters (temperature, error codes, etc.).
Figure 4-6
Signal Interfaces and Power Socket
AC Power Socket
The oven is connected to the AC supply mains via the socket at right in
above,
which accepts a power cord with an IEC standard plug [9].
Fuses
Two 16 A slow-blow fuses mounted on the oven’s main electronic circuit board protect
internal circuitry from the impact of excessive power consumption.
Fuse replacement
The device fuses are not user-serviceable. When the oven exhibits the
typical signs of a blown fuse (no response to pressing the On/Off button,
control panel remains extinguished, no heating operation), call
Customer Service to have the fuses replaced.
Summary of Contents for Heratherm OGS 100
Page 1: ......
Page 6: ...Contents 4 Heratherm Thermo Scientific ...
Page 8: ...List of Figures 2 Heratherm Thermo Scientific ...
Page 34: ...Product Description Work Space Components 4 10 Heratherm Thermo Scientific ...
Page 46: ...Start up Connecting the RS 232 Interface 5 12 Heratherm Thermo Electron LED GmbH ...
Page 62: ...Handling and Control Settings 7 14 Heratherm Thermo Scientific ...
Page 64: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific ...
Page 68: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific ...
Page 84: ...Device Log 15 2 Heratherm Thermo Scientific ...