Start-up
Preparing the Work Space
5-4 Heratherm
Thermo Electron LED GmbH
3. To remove the support rails, pull the retaining spring tab down out of the embossing and
remove the support rail assembly.
Installing and Uninstalling the Rear Air Baffle (OMS Series only)
Heratherm OMS Series ovens are shipped from the factory with an air baffle readily
pre-installed at the back wall of the work space (while OGS Series ovens do not have this
feature). Before the air baffle can be removed from the back wall, the support rails need to be
uninstalled as explained further above.
Figure 5-5
Installing the Rear Air Baffle
1. Losen the two screws [1] that hold the air baffle to the back wall of the work space.
2. Grab the two retaining springs [2] at their tabs and pull them downwards out of the
embossings, then pull off the rear air baffle.
3. For the rear air baffle to install correctly, the two retaining springs [2] must be facing
upwards. Place the air baffle on the lower embossings and tilt it upwards against the back
wall of the work space.
4. Clamp the two retaining springs [2] into the upper embossings.
5. Secure the air baffle to the work space back wall by fastening the two screws at [1].
Installing and Uninstalling the Lateral Air Baffles (OGS Series only)
The following section describes the installation and removal of the lateral air baffles.
Risk of damage to sensor!
To prevent accidental damage, be sure to avoid collision with the sensor
when installing or uninstalling panels in the oven‘s workspace.
Summary of Contents for Heratherm OGS 100
Page 1: ......
Page 6: ...Contents 4 Heratherm Thermo Scientific ...
Page 8: ...List of Figures 2 Heratherm Thermo Scientific ...
Page 34: ...Product Description Work Space Components 4 10 Heratherm Thermo Scientific ...
Page 46: ...Start up Connecting the RS 232 Interface 5 12 Heratherm Thermo Electron LED GmbH ...
Page 62: ...Handling and Control Settings 7 14 Heratherm Thermo Scientific ...
Page 64: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific ...
Page 68: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific ...
Page 84: ...Device Log 15 2 Heratherm Thermo Scientific ...