Cleaning and Disinfection
Wipe / Spray Disinfection
9-2 Heratherm
Thermo Scientific
Preparing the manual wipe/spray disinfection
Alcoholic disinfectants!
Disinfectants having an alcohol content of more than 10% may
form, in combination with air, easily combustible and explosive
gas mixtures.
When using such disinfectants, avoid open flames or exposure to
excessive heat during the entire disinfection process!
Use such disinfectants only in adequately ventilated rooms.
After the disinfectant has been allowed to react, wipe the cleaned
oven components thoroughly dry.
Observe safety regulations to avoid fire and/or explosion hazard
caused by alcohol-containing disinfectants.
Chloride-containing disinfectants!
Chloride-containing disinfectants can corrode stainless steel and
galvanized metal surfaces.
Use only disinfectants that do not affect stainless steel and
galvanized metal surfaces!
Electric shock
Touching live electrical components may cause a lethal electric
shock.
Before connecting the oven to the power supply, check the plug
and power cord for damage. Do not use damaged cables for
connecting the oven to the power source!
Summary of Contents for Heratherm OGS 100
Page 1: ......
Page 6: ...Contents 4 Heratherm Thermo Scientific ...
Page 8: ...List of Figures 2 Heratherm Thermo Scientific ...
Page 34: ...Product Description Work Space Components 4 10 Heratherm Thermo Scientific ...
Page 46: ...Start up Connecting the RS 232 Interface 5 12 Heratherm Thermo Electron LED GmbH ...
Page 62: ...Handling and Control Settings 7 14 Heratherm Thermo Scientific ...
Page 64: ...Shut down Shutting the Oven Down 8 2 Heratherm Thermo Scientific ...
Page 68: ...Cleaning and Disinfection Wipe Spray Disinfection 9 4 Heratherm Thermo Scientific ...
Page 84: ...Device Log 15 2 Heratherm Thermo Scientific ...