![Thermo Scientific HERACELL VIOS 160i User Manual Download Page 51](http://html1.mh-extra.com/html/thermo-scientific/heracell-vios-160i/heracell-vios-160i_user-manual_1104556051.webp)
3
Description of the device
Workspace Components
Thermo Scientific
HERA
CELL VIOS
160i
/
250i
3-21
Heating system
The workspace is heated by a surface heater system. The arrangement of the heating elements ensures
that condensate formation above the water reservoir is prevented.
The outer door of the unit and the perimeter of the door opening are heated as well. The heat radiated
onto the interior glass door/gas-tight screen prevents condensate formation.
The workspace of the device always remains visible, despite high humidity.
Rear panel openings
Figure 3-15.
Rear panel openings
The access port with insert (
Figure 3-15
Ø 42 mm / 1.66 inch) can be sealed with a plug
(1/
Figure 3-15
) and allows for feeding cables, tubes or additional sensors into the workspace of the
unit.
The pressure compensation opening (3/
Figure 3-15
) located directly beneath the access port in the rear
wall of the unit ensures pressure compensation between the workspace of the unit and the operating
room.
1
3
2
Summary of Contents for HERACELL VIOS 160i
Page 8: ...Table of Contents 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 12: ...Figures 4 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 166: ...11 Specifications HERACELL VIOS 250i 11 10 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 196: ...13 Device log 13 2 HERACELL VIOS 160i 250i Thermo Scientific ...