Thermo Scientific
HERA
CELL VIOS
160i
/
250i
2-1
2
Installation of the device
Environmental Conditions
The device must only be operated at locations that meet the particular ambient conditions listed
below:
Requirements:
• Draft-free and dry location.
• The minimal distance to adjacent surfaces must be observed on all sides, see
“Space requirements“
on
page 2-2
.
• The operating room must be equipped with appropriate room ventilation.
• Solid, level, fire-proof surface.
• Vibration-proof substructure (floor stand, lab table) capable of bearing the weight of the device
and of accessories (particularly if several devices are stacked).
• The electrical system of the device has been designed for an operating height of up to 2000 m
above sea level.
• To ensure a constant incubation temperature of 37 °C, the ambient temperature must be within a
range of +18 °C to +34 °C.
• Relative humidity up to 80% (max.).
• No direct exposure to sunlight.
• Devices that produce excessive heat must not be placed near the location of
HERA
CELL VIOS
160i
/
HERA
CELL VIOS
250i
Table of Contents
•
“Environmental Conditions“
on
page 2-1
•
“Room ventilation“
on
page 2-2
•
“Space requirements“
on
page 2-2
•
“Transport“
on
page 2-3
•
“Stacking“
on
page 2-3
•
“Retrofitting/Modifications“
on
page 2-6
Summary of Contents for HERACELL VIOS 160i
Page 8: ...Table of Contents 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 12: ...Figures 4 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 166: ...11 Specifications HERACELL VIOS 250i 11 10 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 196: ...13 Device log 13 2 HERACELL VIOS 160i 250i Thermo Scientific ...