9
Maintenance
Replacing the HEPA-filter
Thermo Scientific
HERA
CELL VIOS
160i
/
250i
9-7
Figure 9-5.
CO
2
calibration procedure
4. Enter the measured value (target value):
The targeting value can be increased or reduced in increments; if you keep the + key or the – key
depressed, the function switches to a rapid increase/reduction; after approx. 3 seconds, another
increase/reduction occurs.
To increase the targeting value:
• Press the + key.
To reduce the set value:
• Press the - key.
5. Accept and save the target value:
• Press the
Enter
key.
• Press the
Save
key.
• The system returns to the main menu. The CO
2
display shows the actual value currently measured
in the workspace.
Replacing the HEPA-filter
The HEPA-filter is mounted under a plastic airbox at the bottom of the rear wall in the workspace.
Steps for replacing the HEPA-filter:
1. Power down the unit, shut off the gas supply and allow the workspace to vent.
2. Take the center and bottom insert out of the workspace.
Note
Excessive CO
2
content:
Excessive CO
2
content after the calibration can be reduced by leaving the device doors open for
approx. 30 seconds.
Value reset:
If the value isn’t changed within the next 30 seconds, the system automatically exits the menu, and
the most recently confirmed value is preserved.
Summary of Contents for HERACELL VIOS 160i
Page 8: ...Table of Contents 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 12: ...Figures 4 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 166: ...11 Specifications HERACELL VIOS 250i 11 10 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 196: ...13 Device log 13 2 HERACELL VIOS 160i 250i Thermo Scientific ...