![Thermo Scientific HERACELL VIOS 160i User Manual Download Page 170](http://html1.mh-extra.com/html/thermo-scientific/heracell-vios-160i/heracell-vios-160i_user-manual_1104556170.webp)
12
Annex: Data communication
Structure of the command sequences for data communication
12-4
HERA
CELL VIOS
160i
/
250i
Thermo Scientific
The interface transmission speed can be selected within the defined baud rates (9,600, 19,200,
38,400, 57,600 Baud) at the touchscreen of the
HERA
CELL VIOS
160i
/
HERA
CELL VIOS
250i
(see
“
Settings / Setup“
on
page 6-17
).
Structure of the command sequences for data
communication
All characters sent and received in the data exchange between a PC and the incubator
HERA
CELL VIOS
160i
/
HERA
CELL VIOS
250i
are ASCII characters that can be displayed on a conventional
terminal.
This ensures that communication can be set up, controlled, and programmed easily.
Description of protocol
Character coding
ASCII characters, capital letters are not allowed.
Reading parameters:
Query:
?:aaaa:bb::cc<CR>
or:
?:aaaa:bb:XXXX:cc<CR>
Response:
!:aaaa:bb:XXXXX:cc<CR>
where:
aaaa = parameter adress
bb =amount of payload in this telegram (00 – ff )
cc = checksum: Inverted XOR of all bytes without
checksum and <CR>
XXXX = bb-bytes user data
Description of the response elements:
aaaa
parameter address
bb
amount of payload in this telegram (00 – ff)
cc
Checksum: Inverted XOR of all bytes without
checksum and <CR>
Example of a software version query (50111927):
Query:
?:0001:00::cc<CR>
Response:
!:0001:08:50111927:cc<CR>
Writing parameters:
Command:
!:aaaa:bb:XXXXX:cc<CR>
Response:
!:aaaa:bb::cc<CR>
where:
aaaa = parameter adress
bb =amount of payload in this telegram (00 – ff )
cc = checksum: Inverted XOR of all bytes without
checksum and <CR>
XXXX = bb-bytes user data
Response with an error message:
Response:
!:aaaa:bb:XX:cc<CR>
Summary of Contents for HERACELL VIOS 160i
Page 8: ...Table of Contents 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 12: ...Figures 4 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 76: ...5 Operation Starting operation 5 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 154: ...9 Maintenance Replacing the Door Seal 9 12 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 166: ...11 Specifications HERACELL VIOS 250i 11 10 HERACELL VIOS 160i 250i Thermo Scientific ...
Page 196: ...13 Device log 13 2 HERACELL VIOS 160i 250i Thermo Scientific ...