Operating Instructions | HERAcell® 150i / 240i
12
Technical data
12
Technical data
12
Technical data
122
50115191 A / 08.2009
HERAcell
®
240i
Description
Unit
Value
Mechanical
External dimensions (W x H x T)
mm
780 x 934 x 834
Interior dimensions (W x H x T)
mm
607 x 670 x 583
Chamber volume
l
approx 238
Shelves (W x T)
mm
560 x 500
Standard quantity
Piece
3
Maximal quantity
Piece
12
Maximal surface load
kg
10 / insertion shelf
Maximal device overall load
kg
30
Weight, without accessories
kg
81
Thermal
Thermal safety devices according to DIN 12880:2007-05
Class 3.1
(overtemperature controller
(TWW) with overtemperature
detection function)
Ambient temperature range
°C
+18...33
Temperature control range
°C
RT + 3 ... 55
Temperature deviation, time (DIN 12880, Part 2)
°C
± 0,1
Temperature deviation, spatial (DIN 12880, Part 2)
at 37 °C *1)
°C
± 0.5
Duration of the auto-start routine, to 37 °C
ambient temperature 20 °C
h
5 ... 10
Heat transfer to environment:
at 37 °C
during contra-con decontamination
kWh/h
kWh/h
0.07
0.25
Humidity
Water quality
resistivity 50 k
Ω
to 1 M
Ω
,
conductivity 1 to 20
μ
S
Liquid quantity:
Incubation operation
contra-con disinfection operation
l
ml
max. 4.5 / min 1.8
350
Constant humidity at 37 °C (high-humidity mode)
Constant humidity at 37 °C (low-humidity mode)
% rH
% rH
approx 93
approx 90
EMC class
B
Others
Sound pressure level (DIN 45 635, Part 1)
dB(A)
< 50
Relative humidity of environment
% rH
max. 80
Location elevation
m ASL
max. 2000
*1) determined on the basis of DIN 12880 for devices with standard equipment. See detailed information in calbration
instructions.