Operating Instructions | HERAcell® 150i / 240i
12
Technical data
12
Technical data
12
Technical data
123
50115191 A / 08.2009
HERAcell
®
240i
Description
Unit
Value
CO
2
gas supply system
Gas purity
%
min. 99.5 or medical quality
Prepressure
bar
min. 0.8 - max. 1
Measuring and control range
% vol.
0... 20
Control deviation, temporal
% vol.
± 0,1
CO
2
measuring cell
Accuracy (absolute)
% CO
2
± 0.3
O
2
gas supply system
Gas purity
%
min. 99.5 or medical quality
Prepressure
bar
min. 0.8 - max. 1
Measuring and control range
% vol.
1... 21 or 5...90
Control deviation, temporal
% vol.
± 0.2
O
2
measuring cell
Accuracy (absolute)
% O
2
± 0.5 (Option: 1.....21 % O
2
)
± 2.0 (Option: 5.....90 % O
2
)
Electrical system
Rated voltage
V
V
V
1/N/PE 230 V, AC (± 10%)
1/N/PE 120 V, AC (± 10%)
1/N/PE 100 V, AC (± 10%)
Rated frequency
Hz
50/60
Interference suppression (DIN VDE 0875)
Interference level N
Type of protection (DIN 40 050)
IP 20
Protection class
l
Overvoltage category (EN 61010)
ll
Pollution severity (EN 61010)
2
Rated current
A
2.8 (230 V, AC)
5.4 (120 V, AC)
6.5 (100 V, AC)
On-site fusing:
Fuse
Circuit breaker
T 10 A
G 16
Rated input
kW
kW
kW
0.64 (230 V, AC)
0.65 (120 V, AC)
0.65 (100 V, AC)