Setting the reminder intervals:
The reminder intervals are integral components of the
alarm and monitoring system of the device control.
For the two essential functions contra-con and auto-
start as well as for routine service work, the user can
set dates that trigger an alarm whenever they occur.
The counting begins at 00:00 hrs of the day on which
the previously set reminder interval has elapsed.
1. To increase the number of days:
>
Press the
+
key [4].
2. To reduce the number of days:
>
Press the
-
key [2].
3. The value change appears in the display [1].
4. To deactivate the reminder interval:
>
Set the value to
O
FF
.
>
Press the
-
key [2].
5. To accept and save the change:
>
Press the
E
NTER
key [3].
>
The system returns to the Reminder Interval
menu.
On the due date, the display shows a reminder
message [5] for the activated reminder interval.
•
contra-con:
Please run contra-con.
•
auto-start:
Please run auto-start. Is displayed after
the contra-con decontamination routine has been
run successfully.
•
Service interval:
Request service. The service
message can be confirmed. In this case, the
Request Service icon is displayed.
After the routines have been run successfully, the
reminder messages are hidden.
Factory settings:
contra-con decontamination routine
90 days
auto-start routine
Off
Service interval
365 days
Fig. 48: Reminder interval setting