Operating Instructions | HERAcell® 150i / 240i
9
Cleaning and disinfection
9
Cleaning and disinfection
105
50115191 A / 08.2009
Optional bottle turning device:
Before starting the contra-con decontamination routine, all
rollers must be removed from the incubator. The electrical
connection sockets must then be sealed with the protecting
caps.
The rack that supports the rollers may be left in place for the
duration of the cycle.
Procedure of a contra-con decontamination routine:
contra-con procedure
1. After cleaning, reinstall the shelf system components into the work
space.
2. Fill the floorpan with 350 ml processed water.
3. Turn the device on using the power switch.
4. Activate and start the decontamination routine.
5. After the decontamination routine has been completed, remove the
remaining water using a sterile cloth.
6. Turn the device off or resume operation.
Hot surfaces!
The handle and the screen of the glass door, the interior panel
of the outer door as well as the surfaces of the shelf system and
of the work space become extremely hot during the contra-con
decontamination routine.
During the routine run or immediately after completion of the
run, always wear safety gloves when touching these surfaces!
Damage to the samples!
During the contra-con decontamination routine, the work
space is heated up to 90 °C.
Make sure that:
• all samples are removed from the work space,
• all accessories are removed from the work space.