Operating Instructions | HERAcell® 150i / 240i
7
Handling and control (iCan™ touchscreen controller)
7
Handling and control (iCan™ touchscreen controller)
7
Handling and control (iCan™ touchscreen controller)
96
50115191 A / 08.2009
Error message overview:
Control
loop
Error message
Cause
Repair
Alarm relay Buzzer
System
Device door open
too long
Doors open for more than 10
minutes.
Close device doors
x
x
Failure display
Communication between
display and main board faulty
Contact Service when failure
is recuring
x
x
Failure EEPROM
main board
EEPROM on main board is
faulty.
Contact Service
x
x
Failure
communication
mainboard
Communication between
display and main board faulty
*1)
Contact Service when failure
is recuring
System - Failure
data logger
Failure during recording data
logger. Device is still
working.
Reset device to cancel failure.
When failure is recuring
contact Service.
x
Failure contra-con
Failure in contra-con routine
Reset device to cancel failure.
When failure is recuring
contact Service.
x
x
System - Power
down during contra-
con
Power failure during contra-
con routine
Reset device and start contra-
con again.
x
x
Failure upon auto-
start
Failure during auto-start
routine
Reset auto-start. When failure
is recuring contact Service.
x
x
System - Safety
circuit active
Temperature signal
plausibility doubtful *2)
Reset device. When failure is
listed several times, contact
Service.
Failure
bottle turning
device
Communication between
bottle turning device and
main board faulty
Contact Service
x
x
Tempe-
rature
Sensor breakage
Detected value is out of
limit.
Contact Service
x
x
Actual value above
Act. val. > nom. val. + 1°C
*3)
Do not exceed ambient
temperature limit
x
x
Actual value below
Act. val. < nom. val. - 1°C
*4)
Contact Service
x
x
Calibration values
too high / too low
Temperature max.
adjustment value exceeded
Contact Service
x
x
Actual value
plausibility
doubtful
Temperature signal
plausibility doubtful
Contact Service
x
x
Failure
communication\
Measuring cell cannot
commuicate with mainboard
Contact Service
x
x
*1) Error message will be displayed, but not entered in error table.
*2) Error message will be displayed, but not entered in error table.
*4) After changing the set value, error time will be set on 159 min., after door opening on 45 min. ( O
2
on 159 min.).
*3) If failure is detected, a special sample protection controlling will be enabled. To indicate the start of this routine,
the icon overtemperature is displayed. The routine will will be disabled after a power reset.
Summary of Contents for Heracell 150i