4. Connect the gas cell outlet to the condensate filter inlet.
5. Install the thermal insulation wrap on the bare tube connections of the HTL and HSL.
Ventilation
This system is designed for vacuum use only. If you are sampling toxic or hazardous gases, the vent
outlet must be connected to a vacuum exhaust vent system designed to safely remove these toxic or
hazardous gases.
If the vent port is blocked or connected to a source of pressure, toxic or hazardous gas could enter
the work area and break the gas cell windows. Never block or plug the vent port or connect it to a
source of pressure.
The pressure relief valve is set to 25 PSI to protect the 2M or 10M gas cell with ZnSe windows. The
relief valve is connected to the vent port.
Thermo Scientific
Modular Gas System User Guide
29
Installation