Installation
Replace gas cell fittings
The gas cell uses 1/4 in. standard Swagelok fittings on the gas inlet and outlets. The fittings may
need to be adapted to 6 mm if you are using metric fittings.
CAUTION
Avoid Leaks. Do not mix 1/4 inch and 6 mm ferrules and nuts. If they
are mixed or if they are attached to the wrong end of the 1/4 inch to
6 mm adapter, the connection will leak.
Figure 3-2: The needle valves on the gas cell must be replaced with a standard fitting
❖
To replace the gas cell fittings
1. Using a 9/16” wrench or an adjustable wrench, loosen the nut closest to the gas cell to remove the
valve.
18