Installation
Step 3: Install the pumps and additional filters
The system may include a vacuum sample pump, a condensate pump, and a water catch filter, which
should also be installed on the rack.
1. Install the vacuum pump.
a. With the gas inlet and outlet valves facing inside the rack, toward the analyzer or heated valve
drawer, place the vacuum pump on the pump bracket with the feet aligned over the screw
holes.
b. To fasten the pump to the rack, reach under the rack and insert each of the four (4) M4 screws
into the holes. Tighten the screws.
2. Install the peristaltic drain pump and water catch.
a. Install the water catch filter behind the heated valve drawer.
16