Installation
The available gas inputs to the system include the following:
l
Heated Sample Line from the sample point to the FTIR Heated Valve Drawer.
l
Nitrogen (N2) used to clean the gas cell and collect a "zero" background sample. You may also
use filtered, compressed air in place of nitrogen. If you are using filted, compressed air, use a filter
that was designed for FTIR use.
l
Optional Span 1 and Span 2 valve positions for span or calibration gas samples.
Gas lines are connected to the FTIR Heated Valve Drawer with Swagelok™ tube fittings.
1. Connect the Heated Sample Line (HSL)
A heated sampling line (HSL) (typically 2 -3 meters in length) connects the sample site to the
FTIR gas cell.Connect the HSL using the standard Swagelok fitting on the rear panel of the
drawer.
Figure 3-4: The Heated Sample Line (HSL) and control box.
a. Fully insert the tubing into theSAMPLE IN fitting.
b. Rotate the nut clockwise until finger tight.
c. Using a wrench, tighten the nut 1¼ turns. Do not over tighten.
d. Connect the round black cable from the HSL to the power control box.
24
Modular Gas System User Guide
Thermo Scientific
Summary of Contents for FTIR
Page 1: ...Modular Gas System USERGUIDE 269 346401B July 2022 ...
Page 15: ... This page intentionally left blank ...
Page 38: ...using the pressure transducer Thermo Scientific Modular Gas System User Guide 33 Installation ...
Page 61: ... This page intentionally left blank ...
Page 79: ... This page intentionally left blank ...
Page 82: ...Eductor Panel Thermo Scientific Modular Gas System User Guide 77 Plumbing diagrams ...