background image

5

 – Troubleshooting Guide  

 

 
 
 

Thermo Scientific 

Product Manual for CarboPac PA200 Columns 

Page 26 of 37 

065734-01 

For Research Use Only. Not for use in diagnostic procedures.

 

5.

 

Troubleshooting Guide 

The purpose of the Troubleshooting Guide is to help you solve operating problems that may arise 

while using Dionex CarboPac columns. For more information on problems that originate with the 

Ion Chromatograph (IC), refer to the Troubleshooting Guide in the appropriate operator’s manual. 

Remember  that  some  of  the  problems  may  be  related  to  parts  of  your  experimental  protocol 

(sample contamination, imprecision during sample transfer, etc.). The following text should help 

you to locate and eliminate problems traceable to the carbohydrate hardware and chemistries. It 

also provides a selection of cleanup and reconditioning procedures that have been found effective 

by many users. 

 

 

For assistance, contact Technical Support for Dionex Products. In the U.S., call 1-800-346-
6390. Outside the U.S., call the nearest Thermo Fisher Scientific office. 

 

5.1 

High Back Pressure 

5.1.1

 

Finding the Source of High System Pressure 

Column  pressure,  (after  subtracting  the  system  pressure)  for  the  Dionex  CarboPac  PA200 

Analytical  Column  should  be  close  to  the  pressure  listed  in  the  QAR  when  using  the  test 

chromatogram conditions.` If a Dionex CarboPac guard and analytical column are both installed, 

column pressure will increase by approximately 20% over the pressure listed in the QAR for the 

column.  If the total system pressure is much higher than expected, it is advisable to determine 

the cause of the high system pressure. 

 

A.

 

Make sure that the pump is set to the correct eluent flow rate. Higher than recommended 

eluent flow rates will cause higher pressure. If necessary, measure the pump flow rate 

by collecting the DI H

2

O eluent for a specified time at operating pressure, and measure 

the collected volume using an analytical balance. This data (weight/time) will give actual 

flow rate. 

B.

 

Determine which part of the system is causing the high pressure. High pressure could 

be due to plugged or constricted tubing, an injection valve with a clogged port, a 

column bed support clogged with particulates, or a clogged detector cell. 

 

To determine which part of the chromatographic system is causing the problem, 

disconnect the pump eluent line from the injection valve and turn the pump on. Watch 

the pressure; it should not exceed 200 psi. (unless a backpressure coil has been 

installed between the pump outlet and the injection valve in which case, first 

disconnect the eluent line from the pump to the backpressure coil).  The pressure with 

the eluent generator connected should be <400 psi. Continue adding system 

components (backpressure coil (if present), injection valve, column(s), and detector) 

one by one, while monitoring the system pressure. The pressure should increase by the 

sum of the measured pressures of the individual guard and analytical columns (see 

product QAR) when the CarboPac Guard and Analytical columns are connected. 

C.

 

Measure the system back pressure by attaching a short piece of new 0.010” tubing in 

place of the column. 

NOTE

!

Summary of Contents for CarboPac PA200

Page 1: ...For Research Use Only Not for use in diagnostic procedures 6 CarboPac PA200 Columns For Dual EGC ICS 6000 065734 Revision 01 January 2018 User Manual ...

Page 2: ...y Not for use in diagnostic procedures Product Manual For CP PA 200 Dual EGC ICS 6000 CarboPac PA200 Guard Column CarboPac PA 200 0 4 x 50 mm Item 302864 CarboPac PA 200 1 x 50 mm Item 302862 CarboPac PA200 Analytical Column CarboPac PA 200 0 4 x 250 mm Item 302863 CarboPac PA 200 1 x 250 mm Item 302861 ...

Page 3: ...ation in this document is for reference purposes only System configurations and specifications in this document supersede all previous information received by the purchaser Thermo Fisher Scientific Inc makes no representations that this document is complete accurate or error free and assumes no responsibility and will not be liable for any errors omissions damage or loss that might result from any...

Page 4: ...ury Indicates a potentially hazardous situation which if not avoided could result in damage to equipment Indicates a potentially hazardous situation which if not avoided may result in minor or moderate injury Also used to identify a situation or practice that may seriously damage the instrument but will not cause injury Indicates information of general interest IMPORTANT Highlights information nec...

Page 5: ...ystem Start up 11 2 6 1 ICS 6000 Dual EG Start Up Procedure for CarboPac PA200 1 mm Column 12 2 6 2 ICS 6000 Dual EG Start Up Procedure for CarboPac PA200 0 4 mm Column 13 2 6 3 Verification of Column Cleanliness 15 3 Operation 16 3 1 Dionex CarboPac PA200 column Operational Parameters 16 3 1 1 The Best Operational Guidelines 16 3 1 2 Initial Check List 16 3 2 Purity Requirements and Preparation o...

Page 6: ...y 29 5 3 2 Shortened Retention Times 30 5 3 3 Loss of Resolution for early eluting peaks 31 5 3 4 Spurious Peaks 31 5 3 5 No Peaks Poor Peak Area Reproducibility or too Small Peak Areas 31 5 3 6 Large Baseline Dip in the Chromatogram 32 5 3 7 Unidentified Peaks Appear with Expected Analyte Peaks 32 5 3 8 Decreased Detection Sensitivity 32 5 3 9 Excessive Gradient Rise 33 5 4 Reconditioning or Repl...

Page 7: ...drophobic polymeric pellicular anion exchange resin stable over the range of pH 0 14 This unique pH stability of the packing material allows the use of eluent compositions that are conducive to anodic oxidation of carbohydrates at gold electrodes Resin Characteristics Particle Size 5 5 µm Pore Size microporous 10 Å Cross linking 55 Ion exchange capacity 1mm 4 0 µeq per column Ion exchange capacity...

Page 8: ...igh pressure applications the use of high pressure fittings is recommended 2 1 The Dionex High Pressure Ion Chromatography Systems A minimum of a Dionex High Pressure Ion Chromatography System HPIC is recommended due to the higher backpressures generated at typical operational flow rates Systems should have the capability to operate up to at least 5000 psi Standard IC systems with an upper limit o...

Page 9: ... below in Figure 1 the flow diagram for a capillary system is shown in Figure 2 Figure 1 Plumbing Schematic for Electrolytic Generation of Potassium Methanesulfonate Eluents Using a Dionex EGC 400 KOH and Dionex EGC 400 MSA Cartridge in an RFIC EG System EGC 400 MSA Cartridge Pump H2O EGC 400 MSA Degasser Injector Guard Column Analytical Column Vacuum Pump KEY 1 Degas Eluent In 2 Degas Eluent Out ...

Page 10: ... wear gloves when handling electrodes Never touch the electrode surface To install a disposable working electrode and reference electrode pH Ag AgCl or PdH refer to Product Manual for Disposable Electrodes Doc No 065040 ICS 5000 Ion Chromatography System Manual Doc No 065342 and User s Compendium for Electrochemical Detection Doc No 065340 In addition the quadruple waveform must be used for carboh...

Page 11: ...he Dionex CarboPac PA200 Analytical Column Retention times will increase by approximately 20 when a guard column is placed in line before the analytical column under isocratic conditions A guard column is utilized to prevent sample contaminants from eluting onto the analytical column It is easier to clean or replace a guard column than an analytical column Replacing the Dionex CarboPac PA200 Guard...

Page 12: ... 400 MSA cartridge for 15 minutes at the flow rate of 1 mL min prime mode F Connect the OUTLET port of the Dionex EGC 400 MSA cartridge to a PEEK backpressure tubing Item 22181 20031 Direct the backpressure tubing to waste G Set the pump flow rate to 0 1 mL min H On the E Panel under Eluent Generator Tab set the Eluent Polarity to Acidic Eluents Set MSAprog to 100 mM Turn on the Dionex EGC power I...

Page 13: ...lumn to waste EE With EGC power OFF turn on the pump and pump DI water at the flow rate of 0 063 mL min for 15 min FF With Eluent Polarity set to Basic Eluents set KMSAprog and KOHprog to the desired concentrations for the applications Turn on the Dionex EGC power for 30 min GG Connect the OUTLET of the column to the CELL IN of the ED cell and direct the CELL OUT to waste HH Confirm that the pH is...

Page 14: ...ity to Basic Eluents Set KOHprog to 100 mM Turn on the Dionex EGC power T Condition the Dionex EGC KOH Capillary for 30 min U On the E Panel turn the Dionex EGC power OFF and keep the pump running at 0 02 mL min for 10 min V Turn off the pump W Remove the PEEK backpressure tubing from the EGC OUT port of the Dionex EGC KOH Capillary cartridge X Connect the EGC OUT port of the Dionex EGC KOH Capill...

Page 15: ...in 2 6 3 Verification of Column Cleanliness Once the columns are equilibrated inject a system suitability standard such as the column s QAR standard to establish the performance of the column at start up This chromatogram can then be referred to when troubleshooting your system Once you obtain the expected chromatographic performance you are ready to proceed to running your application Thermo Scie...

Page 16: ...inutes 3 1 1 2 NEVER Proceed to a next installation step if the previous step has failed Start an installation with any of the check list items below missing Use communal filtration units or filters made of unknown or unsuitable e g cellulose derivatives materials Use MeOH or other organic solvents as rinse fluid in the autosampler Use only water replaced daily or use sampler wash bottles blankete...

Page 17: ... mm Hg that does not deform the plastic eluent container Degas the reservoir for 5 10 minutes while sonicating Cap each bottle and minimize the length of time the bottle is opened to the atmosphere Vacuum filtration through 0 2 μm Nylon filters Fisher Cat 0974046 is a good alternative to vacuum degassing under sonication and is sufficient for the majority of cases On line eluent and EG produced el...

Page 18: ...operated non stop for months at a time If the system has been shut down for more than 1 2 days the system should be re started using the following procedure applied to both analytical and capillary systems 3 5 1 System Re Start Procedure for Dionex CarboPac PA200 1 mm Column A B C Set the flow rate to 0 1 mL min Turn the pump and EGC power on With Eluent Polarity set to Basic Eluents set KMSAprog ...

Page 19: ...t degasser to waste H Set the flow rate to 0 02 mL min and turn the pump on for 10 min Keep EGC power off I Turn the pump off J Connect the ELUENT OUT of the Dionex RFIC Capillary Eluent Degasser Item 22181 60202 to 2 port of the injection valve K Install the CarboPac PA200 0 4 x 50 mm guard column and the CarboPac PA200 0 4 x 250 mm separator column to 3 port of the injection valve Direct the OUT...

Page 20: ...nkage isomerism The disialylated biantennary peaks are eluted before the trisialylated triantennary peaks which are eluted before the tetrasialylated tetraantennary peaks In addition within each grouping the 2 6 isomer is eluted before and well resolved from the 2 3 isomer The Dionex OligoStandard Sialylalted N Linked Alditols Item 043164 contains 25 nmol oligosaccharides purified from bovine fetu...

Page 21: ... 80 min 15 mM KMSA in 136 mM KOH Flow Rate CarboPac PA200 3 mm 0 5 mL min CarboPac PA200 1 mm 0 063 mL min CarboPac PA200 0 4 mm 0 010 mL min Detection CarboPac PA200 3 mm PAD Au on PTFE 2 mil gasket Ag AgCl ref CarboPac PA200 1 mm PAD Au on PTFE 1 mil gasket Ag AgCl ref CarboPac PA200 0 4 mm PAD Au on PTFE 1 mil gasket Ag AgCl ref Samples 50 µmol L fetuin oligosaccharide alditol standard Figure 3...

Page 22: ...m column is for comparison only to demonstrate the comparable performance of the CarboPac PA200 1 mm and 0 4 mm columns with the ease of use Columns CarboPac PA200 3 mm guard separator CarboPac PA200 1 mm guard separator CarboPac PA200 0 4 mm guard separator Gradient CarboPac PA200 3 mm 0 45 min 100 430 mM NaOAc in 100 mM NaOH 45 50 min 430 mM NaOAc in 100 mM NaOH 50 65 min 100 mM NaOAc in 100 mM ...

Page 23: ...tions Thermo Scientific Product Manual for CarboPac PA200 Columns Page 23 of 37 065734 01 For Research Use Only Not for use in diagnostic procedures Figure 4 Inulin Profiles CarboPac PA200 3 mm vs 1 mm vs 0 4 mm ...

Page 24: ...ith manually prepared sodium acetate sodium hydroxide eluents The illustration of the application on the CarboPac PA200 3 mm column is for comparison only to demonstrate the comparable performance of the CarboPac PA200 1 mm and 0 4 mm columns with the ease of use Columns CarboPac PA200 3 mm guard separator CarboPac PA200 1 mm guard separator CarboPac PA200 0 4 mm guard separator Gradient CarboPac ...

Page 25: ... 0 45 0 50 0 55 0 60 0 65 0 70 0 100 50 0 50 100 150 200 250 300 350 400 450 500 550 600 650 700 min nC DP1 DP2 DP3 DP4 DP13 0 4 mm 0 0 5 0 10 0 15 0 20 0 25 0 30 0 35 0 40 0 45 0 50 0 55 0 60 0 20 0 21 0 22 0 23 0 24 0 25 0 26 0 27 0 28 0 29 0 30 0 31 0 32 0 33 0 34 0 35 0 36 0 37 0 min nC DP5 DP6 DP7 DP8 DP9 DP10 DP4 DP11 DP12 Wash step DP13 DP1 DP3 3 mm Zoomed in 0 0 5 0 10 0 15 0 20 0 25 0 30 ...

Page 26: ...R for the column If the total system pressure is much higher than expected it is advisable to determine the cause of the high system pressure A Make sure that the pump is set to the correct eluent flow rate Higher than recommended eluent flow rates will cause higher pressure If necessary measure the pump flow rate by collecting the DI H2O eluent for a specified time at operating pressure and measu...

Page 27: ...y D Place a new bed support assembly provided with each 1 mm column into the end fitting Make sure that the end of the column tube is clean and free of any particulate matter so that it will properly seal against the bed support assembly Drop the bed support assembly into the end fitting making sure that the bed support assembly is centered at the bottom of the end fitting Wrap the end fitting gen...

Page 28: ...m system the Dionex RFIC Capillary EGC MSA Degasser Item 22181 60211 and the Dionex RFIC Capillary Eluent Degasser Item 22181 60202 for the 0 4 mm system D Column contamination Remove the column set from the system first and replace it with a length of 0 001 i d PEEK tubing generating a pressure drop over 2000 psi If the background reading improves after the column is removed from the system go to...

Page 29: ...entrations constant and adjust the gradient time This is the simplest way to compensate for total system differences if resolution is the problem B Change the initial and or final eluent concentration and adjust the gradient time This approach requires more time to develop and more knowledge in methods development work Its advantage is that it allows a method to be tailored for a particular applic...

Page 30: ...n the deionized water used for eluents or components of the sample matrix Be especially careful to make sure that the recommended chemicals are used The deionized water should have a specific resistance of 18 2 megohm cm D Diluting the eluent will improve peak resolution but will also increase the retention times If a 10 dilution of the eluent is not sufficient to obtain the desired peak resolutio...

Page 31: ...mn is used with a weak eluent system the retention times for the analytes will then decrease and spurious inefficient broad peaks may show up at unexpected times Clean the column as indicated in Column Care For assistance contact Technical Support for Dionex Products In the U S call 1 800 346 6390 Outside the U S call the nearest Thermo Fisher Scientific office B The injection valve may need maint...

Page 32: ...he following solutions in the auto sampler and inject in sequence Use 25 µL full loop injections 1 1 M NaOH 2 Deionized water 3 IPA 4 Deionized water 5 1 M HCl 6 Deionized water 5 3 8 Decreased Detection Sensitivity Always confirm the loss of response by performing at least one injection of the system suitability standard mix as described in Section 4 1 This is to make sure that a decreased level ...

Page 33: ...rd analysis This will keep the column conditioned free from carbonate buildup and ready for analysis A Make sure the gradient rise is not caused by the system and or detector cell B Set column temperature to 40 C and wash the guard and column with 100 mM KMSA 100 mM KOH for at least four hours preferably overnight Run a blank gradient at 30 C and if necessary repeat the cleanup with 100 mM KMSA 10...

Page 34: ...ed in the QAR Continue making injections of the test standard until consecutive injections of the standard give reproducible retention times Equilibration is complete when consecutive injections of the standard give reproducible retention times If chromatographic efficiency or resolution is poorer than the QAR see Sections 5 3 Poor Resolution and Section 5 3 1 Loss of Column Efficiency IMPORTANT W...

Page 35: ...umn in series ensure that the guard column is placed after the analytical column in the eluent flow path Contaminants that have accumulated on the guard column can be eluted onto the analytical column and irreversibly damage it If in doubt clean each column separately B Wash the Dionex CarboPac PA200 with deionized water 18 2 megohm cm for about 30 minutes and then clean with 200 mM MSA for one to...

Page 36: ... Quality Assurance Report QAR Thermo Scientific Product Manual for CarboPac PA200 Columns Page 36 of 37 065734 01 For Research Use Only Not for use in diagnostic procedures Appendix B Quality Assurance Report QAR ...

Page 37: ...Appendix B Quality Assurance Report QAR Thermo Scientific Product Manual for CarboPac PA200 Columns Page 37 of 37 065734 01 For Research Use Only Not for use in diagnostic procedures ...

Reviews: